03 ตุลาคม 2567
ผู้ชม 75 ผู้ชม
G3030-1S Industrial Micro Syringe Pump, newly upgraded.Equipped with a high-precision feedback system, significantly enhancing detection sensitivity and operational stability. Compatible with various valve specifications and has a lifespan of over one million cycles, effectively reducing maintenance costs. Meticulously designed for needs such as pipetting, diluting, and dispensing, it is widely used in equipment support systems like environmental detectors, spray coating machines, and pipetting workstations, meeting high-performance fluid handling requirements in both industrial and research settings.
Max: 1 Channel
Features
1. High-Precision Feedback System
Even in complex environments, this system maintains excellent detection sensitivity and operational stability, ensuring high precision and
consistency in fluid handling.
2. Versatile and Durable
Equipped with intelligent limit protection, it accurately monitors the pump’s operating range to prevent over-limit operations, providing
comprehensive safety assurance.
3. Easy Installation
The manual pull-down installation base allows for quick syringe replacement. Its foolproof design simplifies installation and improves
operational efficiency.
4. Convenient Setup and Quick Upgrade
DIP switch settings enable fast valve switching and offline operation, making the process smoother. It also supports online upgrades, enhancing
flexibility and adaptability.
5. Command Set Compatibility
The unique command set is perfectly compatible with mainstream market products, offering broader application support.
6. Stable Connection
The gold-plated interface ensures a stable and reliable connection.
7. Intelligent Limit Protection
This system precisely monitors the pump’s operating range to prevent over-limit operations, providing safety assurance.
8. High-Performance Components
Durable lead screws and powerful motors increase the equipment’s lifespan, precision, and efficiency, while low-noise operation ensures a
comfortable working experience.
9. Flexible Configuration
It can be used independently or in a cascade with up to 15 pumps, adapting to various fluid handling needs.
Parameters
Linear speed: 0.005mm/s~25mm/s
Running time: 1.2s~100min
Flow accuracy: ≤±1%
Pusher advance per microstep: 0.01mm/step
Channel: 1
Power supply: DC24V±10%,electric current≤1.5A
Syringe: 50uL,100uL,250uL,500uL,1mL,2.5mL,5mL
Rated stroke: 30mm (3000 step(standard/24000 step(microstep))
Valve fitting: 1/4~28UNF internal thread
Piston driving force: Stepper motor drive, orthogonal encoding detection position
Applicable valve: 3-way valve,4-way valve,6-way valve,9-way valve
Valve displacement time: ≤280ms
Materials in contact with fluids: Borosilicate glass, Kel~F and Teflon
Communication interface: RS232/RS485,9600/38400 baud rate optional
Parameter setting: 7 digit dial switch
Signal input: 2 gates TTL Level signal input
Signal output: 3 gates TTL level signal outpump
Shell material: Aluminum alloy
Programming function: Ramp, cut-off speed, gap compensation, infuse speed, circulation, real-time termination of movement and delay, error detection, valve rotation selection, storage of 15 groups of operating modes
Working environment: Temperature 15℃~40℃,relative humidity<80%
Dimension(L×W×H): 134×44.5×127(mm)
Weight: 1.23kg
Above flow parameters are obtained by transfer pure water under normal temperature and pressure, in actually using it is effected by specific factors such as pressure, medium etc. Above for reference only.
Connector and Tubeline
Thread inverted taper joint
Specification | 1/4-28UNF |
Material | PTFE |
Transmission line
Specification | 0.5×1.6mm 1.0×1.6mm 1.6×3.2mm |
Material | PTFE |
Syringe Flow Rate Patameters
Syringe Size
|
Min Flow Rate(mL/min)
|
Max Flow Rate(mL/min)
|
50μl
|
0.0005
|
2.5
|
100μl
|
0.0010
|
5
|
250μl
|
0.0025
|
12.5
|
500μl
|
0.005
|
25
|
1ml
|
0.010
|
50
|
2.5ml
|
0.025
|
125
|
5ml
|
0.05
|
250
|
Above flow parameters are obtained by using silicone tube to transfer pure water under normal temperature and pressure, in actually it is effected by specific factors such as pressure, medium, etc. Above for reference only.