05 ตุลาคม 2566
ผู้ชม 1645 ผู้ชม
Three-Zone Furnace Chamber Control
The temperature is measured by thermocouples positioned outside of the working tube, one in the middle and two on the sides. The outer zones are controlled with a setpoint-offset in relation to the middle zone. This allows the heat loss at the ends of the tube to be compensated to ensure an extended zone of constant temperature (+/- 5 K). Furnace Chamber Control with temperature measurement in furnace chamber outside the working tube.
Extension Package for Furnace Chamber Control
Furnace Chamber Control Only the furnace chamber temperature is measured and controlled. Regulation is carried out slowly to avoid out-of-range values. As the charge temperature is not measured and controlled, it may vary a few degrees from the chamber temperature. Charge ControlIf the charge control is switched on, both the charge temperature and furnace chamber temperature are measured. By setting different parameters the heat-up and cooling processes can be individually adapted. This results in a more precise temperature control at the charge. |
PDF Download |
ขอใบเสนอราคา |
---|
Reference Customer (กดที่โลโก้ เพื่อดูเว็บไซต์ของลูกค้าที่สั่งซื้อสินค้ากับทางบริษัท) | ||||
อยู่ในระหว่างอัพเดทข้อมูล... |